ORIGINAL ARTICLE |
|
|
|
|
Year : 2011 | Volume
: 15
| Issue : 2 | Page : 161-167 |
|
A study to evaluate the efficacy of xylene-free hematoxylin and eosin staining procedure as compared to the conventional hematoxylin and eosin staining: An experimental study
Madhuri R Ankle, Priya S Joshi
Department of Oral Pathology and Microbiology, Vasantdada Patil Dental College and Hospital, Kavalapur, Sangli, India
Date of Web Publication | 5-Sep-2011 |
Correspondence Address: Madhuri R Ankle Department of Oral Pathology and Microbiology, Vasantdada Patil Dental College and Hospital, Kavalapur, Sangli, Maharashtra India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/0973-029X.84482
Abstract | | |
Context: Use of diluted dish washing solution (DWS) has been experimented successfully as a substitute for xylene to deparaffinize tissue sections during hematoxylin and eosin (H and E) staining. Aims: (1) Test the hypothesis that xylene- and methanol-free sections (XMF) deparaffinized with diluted DWS are better than or at par with conventional H and E sections. (2) To compare the efficacy of xylene-free sections with the conventional H and E sections. Settings and Design: Single blinded experimental study. Materials and Methods: Sixty paraffin blocks were considered. One section was stained with conventional H and E method (Group A) and the other with XMF H and E (Group B). Slides were scored for parameters; nuclear staining, cytoplasmic staining (adequate = score1, inadequate = score0), uniformity, clarity, crispness (present = score1, absent = score0). Score >/= 2 was inadequate for diagnosis and 3-5 was adequate for diagnosis. Statistical analysis used: Z test. Results: Adequate nuclear staining, 96.66% sections in group A and 98.33% in Group B (Z = 0.59, P>0.05); adequate cytoplasmic staining, 93.33% in group A and 83.33% in Group B (Z = 1.97, P<0.05); uniform staining, 70% in group A, 50% in group B (Z = 1.94, P<0.05), clarity present in 85% of group A, 88.33% of group B sections (Z = 0.27, P>0.05), crisp staining in 76.66% in group A and 83.33% in Group B (Z = 1.98, P<0.05), 88.33% Group A sections stained adequately for diagnosis as compared with 90% in Group B (Z = 0.17, P>0.05). Conclusion: Xylene- and methanol-free H and E staining is a better alternative to the conventional H and E staining procedure.
Keywords: Deparaffinization, eosin, hematoxylin, liquid dish washing soap, xylene-free staining
How to cite this article: Ankle MR, Joshi PS. A study to evaluate the efficacy of xylene-free hematoxylin and eosin staining procedure as compared to the conventional hematoxylin and eosin staining: An experimental study. J Oral Maxillofac Pathol 2011;15:161-7 |
How to cite this URL: Ankle MR, Joshi PS. A study to evaluate the efficacy of xylene-free hematoxylin and eosin staining procedure as compared to the conventional hematoxylin and eosin staining: An experimental study. J Oral Maxillofac Pathol [serial online] 2011 [cited 2023 Mar 24];15:161-7. Available from: https://www.jomfp.in/text.asp?2011/15/2/161/84482 |
Introduction | |  |
The hematoxylin and eosin (H and E) stain forms the backbone of daily pathological diagnostic work. The basic material for the bulk of daily diagnostic work in a pathology laboratory is the paraffin section, usually stained with H and E. [1] H and E is a universal stain and is a primary contrast method in medical diagnosis of biopsy specimen. H and E staining is remarkably robust. It is used to discriminate between a broad range of cytoplasmic, nuclear and extracellular matrix features. [2] This staining procedure has remained unchanged for over 150 years. [1] Apart from hematoxylin and eosin, the components in the H and E staining procedure are the xylene and graded alcohols. These chemicals are used to carry out the intermediate steps of rehydration and dehydration of tissue sections during the staining. The lacunae however that continue to persist in this age-old procedure are the cost containment, toxicity, problem of disposal of the hazardous chemicals; xylene and methanol, and a polluted working environment. [1]
Exposure to xylene in a laboratory occurs during tissue processing, deparaffinization of tissue sections, cover slipping, cleaning tissue processors and recycling. Exposure to methanol occurs during tissue processing and dewaxing the sections before staining. [3]
In the quest to eliminate the use of xylene from the laboratory, numerous substitute chemicals like limonene reagents, aliphatic hydrocarbons, aromatic hydrocarbons, vegetable oils, olive oil, and mineral oil substitutes have been utilized. However, these chemicals were used to substitute xylene as a clearing agent during routine processing, while the exposure and handling of xylene is maximum during dewaxing of the tissue sections. [3]
Basic aim in any field of life sciences is to utilize eco-friendly chemicals which are nontoxic, less biohazardous, and are economical. Liquid dish washing soap (DWS) forms a part of the day-to-day household job. It is a detergent liquid used to clean the greasy utensils in a kitchen.
The innovative concept of using liquid dish washing detergent to dewax the tissue sections by eliminating both xylene as well as alcohol from the staining task was experimented by Falkeholm et al.[1]
Aim
Therefore, an experimental, cross-sectional study was carried out,
- To test the hypothesis that xylene- and methanol-free (XMF) sections deparaffinized with diluted DWS are better than or at par with the conventional H and E-stained sections.
- To compare the efficacy of XMF sections with the conventional H and E sections in producing adequate H and E staining.
Materials and Methods | |  |
The study was carried out in the Department of Oral Pathology and Microbiology. Sixty paraffin blocks of the routine biopsy specimen were retrieved from the Department.
Two paraffin sections of 4 microns each were cut from each of the 60 paraffin blocks of routinely processed tissue specimen. One was stained with the conventional (routine) H and E staining method. The standard time period of the H and E staining procedure ranges from 70 to 75 minutes [Table 1] and the other section was stained with the XMF H and E staining method. This H and E staining procedure is completed in 30 to 35minutes [Table 2]. Thus, a total of 120 sections were studied.
Sixty tissue sections stained with conventional H and E--(Group A) and 60 sections stained with the XMF method--(Group B) were coded. A randomized mix of 120 sections gave 60 matched pairs. Each section was scored and analyzed by a single oral pathologist who was blinded.
H and E-stained sections were graded based on the parameters of nuclear staining (Adequate = score 1, Inadequate = 0), cytoplasmic staining (Adequate = score 1, Inadequate = score 0), clarity of staining (Present = score 1, absent = score 0), uniformity of staining (Present = score 1, absent = score 0), and crispness of staining (Present = score 1, absent = score 0). The scores for each slide were totalled. A score of ≥ 2 was graded as inadequate for diagnosis, slides with score 3-5 were assigned as adequate for diagnosis.
Results | |  |
Adequate nuclear staining was noted in 96.66% of sections of group A as compared with 98.33% in Group B (Z = 0.59, P>0.05) [Table 3], adequate cytoplasmic staining was noted in 93.33% of sections in group A as compared with 83.33% in Group B (Z = 1.97, P<0.05) [Table 4], clarity was present in 85% of group A and in 88.33% of group B sections (Z = 0.27, P>0.05) [Table 5], uniform staining was present in 70% of group A and in 50% of group B (Z = 1.94, P<0.05) [Table 6], and a crisp staining was seen in 76.66% of sections in group A as compared with 83.33% in Group B (Z = 1.98, P<0.05) [Table 7]. The staining was found to be adequate for diagnosis in 88.33% of Group A sections as compared with 90% in Group B (Z = 0.17, P>0.05) [Table 8]. | Table 8: Scores for the adequacy for diagnosis of the stained sections in group A and B
Click here to view |
Discussion | |  |
Xylene forms an inseparable part of a pathology laboratory. The historical use of xylene in the histology laboratory is an example of a failed substitution. Starting as the safest alternative to dangerous chemicals such as aniline oil, benzene, chloroform, dioxane, and toluene in the 1950s, by the 1970, there were great concerns about its safety. Toxic effects of xylene include acute neurotoxicity, cardiac and kidney injuries, fatal blood dyscrasias, skin erythema, drying, scaling, and secondary infection. [3],[4] Methanol is also known to be toxic. The knowledge of xylene toxicity resulted in 41% of US histology laboratories to use xylene substitutes. Countries like US have provisions like OSHA (Occupational Safety and Health Administration) and ACGIH (American Conference of Governmental Industrial Hygienist). These bodies have established standards for biological exposure limits, for monitoring exposure and managing the disposal and recycling of xylene. [3] However, in most of the developing countries, especially in India, we find no such provisions. The institutional and private pathology laboratories in developing countries have neither any mechanism for monitoring the exposure nor any standardized methods for disposal of xylene.
Therefore, any technique that minimizes the use of xylene by using non-biohazardous substitutes, reduces staining time, and does not compromise the staining quality will be very valuable for diagnostic reasons as well as for maintaining a healthy laboratory environment, thereby minimizing the risk to the laboratory personell. Thus, in an effort to improve the working conditions in a histopathology laboratory, we investigated the less toxic, cheap, easily available diluted liquid DWS as a deparaffinizing agent for H and E staining method.
The results showed that of the 60 sections studied, 98.33% of the Group B (XMF) slides showed adequate nuclear staining as compared with 96.66% of the conventional H and E [Table 3]. The difference was not statistically significant (Z = 0.59, P>0.05) suggesting that there was no difference in the two staining methods in producing adequate nuclear staining [Graph 1] [Figure 1]. The nuclear staining in the xylene-free staining method was done using Mayer's hematoxylin. Commercially available Mayer's hematoxylin (Yucca Diagnostics, Kolhapur, Maharashtra, India) was used. It is a progressive stain. [5] Progressive staining can be easily controlled, with less chances of overstaining. There is no differentiation required (thereby saving time) unlike with the regressive hematoxylin stains like Harri's and Delafield's hematoxylin. Moreover, incomplete differentiation following regressive staining can result in inadequate cytoplasmic staining due to the binding of aluminium hematin in the cytoplasm. [6] The degree of crispness of nuclear staining with Mayer's was found to be nearly equivalent to Harri's hematoxylin. | Figure 1: Photomicrograph showing adequacy and clarity of nuclear and cytoplasmic staining (a) Routine H and E staining, (H and E, 100×), (b) Xylene- and alcohol-free H and E staining, (H and E, 100×)
Click here to view | 
83.33% of XF sections had adequately stained cytoplasm as compared with the 93.33% of the conventional H and E sections (Z = 1.97, P<0.05) [Table 4]. A statistically significant downgradation of cytoplasmic staining was noted in the xylene-free H and E-stained sections [Graph 2]. Commercially available water-soluble 1% Eosin Y (Yucca Diagnostics, Kolhapur, Maharashtra, India) was used with the XMF staining. Eosin Y brought out the maximum details of the cytoplasm in 50 of 60 XMF sections studied [Figure 1]. Ten sections showed a deteriorated cytoplasmic stain. The sections appeared bluish [Figure 2]a. Eosin is an acidic dye with optimum staining occurring at a pH of 5.2 to 5.4. It will rinse out in any alkaline solution. [7] The pH of eosin was found to be adequate. The other sources of alkalinity was the tap water wash used before and after the eosin staining step. Therefore, the tap water was analyzed for the pH. It had a pH of 7.01. Lithium carbonate was another source of alkalinity. Increase of alkalinity had probably resulted in eosin being washed out, resulting in a light staining. The pH of tap water was found to be adequate for bluing and hence lithium carbonate was eliminated. This resulted in overcoming this problem. | Figure 2: Photomicrograph showing (a) deterioration of cytoplasmic staining in (10 of 60) xylene- and alcohol-free sections. Xylene- and alcohol-free H and E, 40×, (b) out-of-focus areas in the xylene- and alcohol-free H and E-stained sections, (H and E, 10×)
Click here to view |

The clarity with XMF sections was 88.33% as compared with 85% of clarity in conventional (routine) H and E staining [Table 5]. No statistically significant difference was noted in the two staining methods followed, suggesting that the XMF H and E are at par with conventional H and E in producing clarity in staining [Graph 3]. Xylene is a known carcinogen with numerous biotoxic effects. Histology, however, continues to use large quantities of organic solvents and remains a major producer of organic solvent effluent. [3] The diluted liquid DWS is an extremely cheap, nontoxic substitute for xylene. It is readily available in any stationery stores. The liquid DWS is composed of sodium laureth sulfate, sodium dodecyl benzene sulfonate, cocamidopropyl betaine, and nonionic surfactants. These components are anionic surfactants commonly used in detergent soaps and shampoos. [8] These chemicals form a part of the products which are used daily. Their concentration in these products is already well monitored by the manufacturing companies. Moreover, we are diluting only 25 ml of the liquid DWS in 1 500 ml of distilled water. Thus, there are meager chances of this product being toxic to the laboratory personnel.
The uniformity of the stained section was downgraded significantly in the XMF sections as compared with the conventional H and E sections (70%) (Z = 1.94, P<0.05). This was shown in the results which showed only 50% of the XMF sections to be uniformly stained [Table 6] [Graph 4]. Out-of-focus areas seen in the section compromised the uniformity. This was a consistent problem encountered as seen in the [Figure 2]b. Out-of-focus areas can be due to the reasons like tear or rip of section, introduction of extraneous tissue, unclean blade, dirty microscopic lenses, thick section, and moisture on coverslip. [9] All of these reasons were ruled out. Careful scrutiny and retrospective analysis of the staining procedure led to an important conclusion that the xylene-free staining procedure is highly temperature sensitive. The diluted 1.7% liquid DWS-I and II and the distilled water I and II had to be strictly maintained at 90°C. A slight drop in the temperature failed to deparaffinize the sections completely. This resulted in microscopic amounts of residual wax on the tissue sections that resulted in out-of-focus areas.
83.33% of the XMF sections revealed a crisp staining as compared with 76.66% of the conventional H and E stain [Table 7]. The XMF sections showed a significant upgradation in crispness as compared with the conventional H and E stain (Z = 1.98, P<0.05) [Graph 5]. The combination of 1.7% liquid DWS along with water-soluble Mayer's hematoxylin and eosin Y brings about an ideal degree of crispness. Deparaffinization with diluted liquid DWS is achieved in only 4 minutes unlike 20 minutes required for deparaffinization with xylene and alcohol in the routine H and E staining. This saves time and simplifies the staining procedure [Figure 3] and [Figure 4]. | Figure 3: Photomicrograph showing the crisp staining of epithelium, connective tissues, and glandular tissue. (a) Epithelium and connective tissue stained with routine H and E staining, (H and E, 10×), (b) Epithelium and connective tissue stained with xylene- and alcohol-free H and E staining, (H and E, 10×), (c) Glandular stained with routine H and E staining, (H and E, 40×), (d) Glandular tissue stained with xylene- and alcohol-free H and E staining, (H and E, 40×)
Click here to view |
 | Figure 4: Photomicrograph showing the crisp staining of muscle and bone tissues. (a) Muscle tissue stained with routine H and E staining, (H and E, 40×), (b) Muscle tissue stained with xylene- and alcohol-free H and E staining, (H and E, 40×), (c) Bone tissue stained with routine H and E staining, (H and E, 40×), (d) Bone tissue stained with xylene- and alcohol-free H and E staining, (H and E, 40×)
Click here to view |

When the scores were totaled, 90% of the XMF H and E-stained sections were found to be adequate for diagnosis as compared with 88.33% of the slides stained with routine H and E (Z = 0.17, P>0.05) [Table 8] [Graph 6]. Falkeholm et al.[1] found that the xylene-free sections were ranked as good as or better than their conventional counterparts in 74% of the comparisons, and poorer in 26%. Beusa and Peshkov [3] demonstrated that 1.7% dishwasher soap aqueous solution at 90°C before staining and overdrying the sections after staining will eliminate xylene from the staining task. The advantages of the xylene- and alcohol-free H and E staining procedure are listed in [Table 9].
Conclusion | |  |
Thus, the present study shows that XMF staining procedure carried out using a simple diluted liquid dish washing soap solution is at par with the conventional H and E procedure. It produces a quality staining with sufficient clarity and a crisp nuclear and cytoplasmic staining. It also has added advantages of being nontoxic, economical, nonflammable, nonhazardous, no problem of disposal, reduces staining time, and is easy to handle.
Scope for The Study | |  |
The xylene- and alcohol-free H and E-stained sections need to be evaluated after a time period of 1 year or so in order to see for the stability in the staining of tissue section and to see that the sections have not faded.
Acknowledgments | |  |
We are thankful to Dr. Hemant Pawar, Assistant Professor, Department of Preventive & Community Medicine, Pravara Institute of Medical Sciences, Loni, for the statistical work-up. We are also thankful to Dr. Someshwar Golgire, postgraduate in the Department of Oral Pathology, VPDC, Kavalapur, Sangli, and to Mr. Mohan Jagtap, Laboratory Technician in our Department for their assistance during the study.
References | |  |
1. | Falkeholm L, Grant CA, Magnusson A, Möller E. Xylene-Free Method for Histological Preparation: A Multicentre Evaluation. Lab Invest 2001;81:1213-21.  |
2. | Q Imaging Camera Application Notes. H and E stain tissue documentation. [Internet] Available from http://www.qimaging.com/support/pdfs/he_technote.pdf [Last cited on 2010 Dec 24].  |
3. | Buesa RJ, Peshkov MV. Histology without xylene. Ann Diagn Pathol 2009;13:246-56.  [PUBMED] [FULLTEXT] |
4. | Kandyala R, Raghavendra SP, Rajasekharan ST. Xylene: An overview of its health hazards and preventive measures. J Oral Maxillofac Pathol 2010;14:1-5.  [PUBMED] |
5. | Alan Stevens and Ian Wilson. Chapter 6. The Haematoxylins and Eosin. In: Bancroft JD, Stevens A, editors. Theory and Practice of Histological Techniques. 4th Ed. New York: Churchill-Livingstone; 1996. p. 99-112.  |
6. | Gill G. H&E staining: Oversights and Insights. [Internet] Available from: www.dako.com/08066_12may10_webchapter13.pdf [Last cited on 2010 Dec 24].  |
7. | Ellis R. Problems in Histopathological Technique. [Internet]. Available from: www.ihcworld.com/royellis/problems/problem25.htm [Last cited on 2010 Dec 24].  |
8. | Available from: Internet source: en.wikipedia.org/wiki/Detergent [Last accessed on 2010 Dec 24].  |
9. | Hamill JR, Spencer S, Morgan. On Atlas of Mohs and Frozen Section Cutaneous Pathology [Internet]. New York: Springer Link; 2009. Chapter 20, Histotechnique and Staining Troubleshooting. Available from: http://www.books.google.co.in/books. [Last cited on 2010 Dec 24].  |
[Figure 1], [Figure 2], [Figure 3], [Figure 4]
[Table 1], [Table 2], [Table 3], [Table 4], [Table 5], [Table 6], [Table 7], [Table 8], [Table 9]
This article has been cited by | 1 |
Label-free detection of invasive micropapillary carcinoma of the breast using multiphoton microscopy |
|
| Jiajia He, Deyong Kang, Tingfeng Shen, Liqin Zheng, Zhenlin Zhan, Gangqin Xi, Wenjiao Ren, Zhong Chen, Lida Qiu, Shuoyu Xu, Lianhuang Li, Jianxin Chen | | Journal of Biophotonics. 2022; | | [Pubmed] | [DOI] | | 2 |
Comparing the efficacy of xylene and kerosene oil as a clearing agent during processing and staining in histopathology |
|
| Manbeer Singh, Amit Kumar Singh, Jappreet Kaur | | Indian Journal of Pathology and Oncology. 2022; 9(3): 253 | | [Pubmed] | [DOI] | | 3 |
Utilization Dishwashing Soap as a Subtitute of Xylol in the Deparaffinization process of Hematoxylin-Eosin Dye: Review Article |
|
| Mamay Mamay, Gina Nafsa Mutmaina, Ina Aflaha Nurahma | | Medicra (Journal of Medical Laboratory Science/Technology). 2022; 5(1): 47 | | [Pubmed] | [DOI] | | 4 |
Interplay of collagen and mast cells in periapical granulomas and periapical cysts: a comparative polarizing microscopic and immunohistochemical study |
|
| Deepty Bansal, Mala Kamboj, Anjali Narwal, Anju Devi, Nisha Marwah | | Restorative Dentistry & Endodontics. 2022; 47 | | [Pubmed] | [DOI] | | 5 |
A comparative study of efficacy of coconut oil, lemon water and dishwashing liquid as surrogates to xylene |
|
| Ayeesha Thajudeen, Sudha Srinivasan, GeethaPriya Govindarajan, Akashavanan Shanmugam | | Environmental Analysis Health and Toxicology. 2022; 37(3): e2022026 | | [Pubmed] | [DOI] | | 6 |
Understanding the sources of errors in ex vivo Hsp90 molecular imaging for rapid-on-site breast cancer diagnosis |
|
| Roujia Wang, Daniel A. Alvarez, Brian T. Crouch, Aditi Pilani, Christopher Lam, Caigang Zhu, Philip Hughes, David Katz, Timothy Haystead, Nirmala Ramanujam | | Biomedical Optics Express. 2021; 12(4): 2299 | | [Pubmed] | [DOI] | | 7 |
A COMPARATIVE STUDY OF THE EFFICACY OF
CEDARWOOD OIL, COCONUT OIL AND DISH
WASH LIQUID AS ALTERNATIVES TO XYLENE AS
DEPARAFFINIZING AGENTS |
|
| Sithara K, Ganesh Prasad B, Dinkar Desai | | PARIPEX INDIAN JOURNAL OF RESEARCH. 2021; : 33 | | [Pubmed] | [DOI] | | 8 |
Enhancement of immune responses by co-stimulation of TLR3 - TLR7 agonists as a potential therapeutics against rabies in mouse model |
|
| Firouzeh Farahtaj, Alireza Gholami, Mohammad Sadeq Khosravy, Safoora Gharibzadeh, Hamid Mahmoudzadeh Niknam, Amir Ghaemi | | Microbial Pathogenesis. 2021; 157: 104971 | | [Pubmed] | [DOI] | | 9 |
The protective effect of shrimp cooked in different methods on high-cholesterol- induced fatty liver in rats |
|
| Nora A. AlFaris, Ghedeir M. Alshammari, Jozaa Z. AlTamimi, Lujain A. AlMousa, Nora M. AlKehayez, Dalal H. Aljabryn, Reham I. Alagal, Mohammed A. Yahya | | Saudi Journal of Biological Sciences. 2021; 28(1): 170 | | [Pubmed] | [DOI] | | 10 |
Biofriendly substitutes for xylene in deparaffinization |
|
| Viswanathan Prema, Harikrishnan Prasad, KenniyanKumar Srichinthu, SingaraveluSuresh Kumar, Krishnan Rajkumar, Chinnannan Marudhamani | | Journal of Pharmacy And Bioallied Sciences. 2020; 12(5): 623 | | [Pubmed] | [DOI] | | 11 |
Microscopic Validation of Macroscopic In Vivo Images Enabled by Same-Slide Optical and Nuclear Fusion |
|
| Kazumasa Inoue, Summer L. Gibbs, Fangbing Liu, Jeong Heon Lee, Yang Xie, Yoshitomo Ashitate, Hirofumi Fujii, John V. Frangioni, Hak Soo Choi | | Journal of Nuclear Medicine. 2014; 55(11): 1899 | | [Pubmed] | [DOI] | | 12 |
Efficacy of 1.5% Dish Washing Solution and 95% Lemon Water in Substituting Perilous Xylene as a Deparaffinizing Agent for Routine H and E Staining Procedure: A Short Study |
|
| Anuradha Ananthaneni,Srilekha Namala,Vijay Srinivas Guduru,V. V. S. Ramprasad,Sabitha Devi Ramisetty,Urmila Udayashankar,Kiran Kumar Naik | | Scientifica. 2014; 2014: 1 | | [Pubmed] | [DOI] | | 13 |
Health hazards of xylene: A literature review |
|
| Rajan, S.T., Malathi, N. | | Journal of Clinical and Diagnostic Research. 2014; 8(2): 217-274 | | [Pubmed] | | 14 |
Mineral oil-A biofriendly substitute for xylene in deparaffinization: A novel method |
|
| Premalatha, B.R., Patil, S., Rao, R.S., Indu, M. | | Journal of Contemporary Dental Practice. 2013; 14(2): 281-286 | | [Pubmed] | | 15 |
Biosafe alternative to xylene: A comparative study |
|
| Negi, A., Puri, A., Gupta, R.,Nangia, R., Sachdeva, A. | | Journal of Oral and Maxillofacial Pathology. 2013; 17(3): 363-366 | | [Pubmed] | | 16 |
Conventional xylene and xylene-free methods for routine histopathological preparation of tissue sections |
|
| Metgud, R. and Astekar, M.S. and Soni, A. and Naik, S. and Vanishree, M. | | Biotechnic and Histochemistry. 2013; 88(5): 235-241 | | [Pubmed] | | 17 |
The application of heated detergent dewaxing and rehydration to techniques for the demonstration of fungi: A comparison to routine xylene-alcohol dewaxing |
|
| Henwood, A.F. and Prasad, L. and Bourke, V.-M. | | Journal of Histotechnology. 2013; 36(2): 45-50 | | [Pubmed] | |
|
 |
|
|
|
|